EDAN F9 Express Fetal Monitor

The Edan F9 Express Fetal/Maternal Monitor provides solutions for all areas of fetal monitoring, ranging from antepartum, intrapartum, and postpartum clinical applications.

Integrated monitoring of twins FHR, uterine activity, fetal movement, intrauterine pressure and direct ECG combined with maternal NIBP, SpO2, ECG, and TEMP, provides cost effective and flexible solutions for all fetal monitoring needs, without compromising quality, precision, performance, or ergonomics.

More Information

Features:

  • Sensitive waterproof FHR transducer
  • Flat surface design and waterproof TOCO transducer
  • Fetal and maternal parameters
  • Monitoring, storage and playback
  • Three monitoring interfaces
  • Integrated thermal printer
  • Compatible with 150/152 mm universal recording paper
  • Signals Overlap Verification to separate twins FHR
  • Continuous post-delivery monitoring with maternal parameters: NIBP, SpO2, HR, ECG, TEMP
  • Optional telemetry transducer system
  • 24 hours waveforms playback
  • Optional built-in wireless module
  • Software for data transmission to PC
  • 1, 2, 3 cm/minute real-time printing speeds
  • Quick printing of stored waveform
  • Rechargeable lithium-ion battery delivers 4 hours continuous monitoring
  • Transducer mounting rack or wall mounting rack
  • Handle for portability

Specifications:

  • Basic Parameters: FHR, TOCO, AFM
  • Internal Parameters: IUP, DECG (optional)
  • Maternal Parameters: MECG, NIBP, MSpO2
  • Display: 12.1″ (30.7 cm) colour TFT-LCD touchscreen
  • Display Resolution: 800 x 600
  • Dimensions: 347 mm x 330 mm x 126 mm
  • Weight: 6 kg
  • Battery: 14.8 V rechargeable lithium-Ion battery
  • Ultrasound:
    • Technique: Ultrasound Pulse Doppler with correlation
    • Pulse Repetition Rate: 2 KHz
    • Ultrasound Frequency: 1.0 MHz + 10%
    • FHR Range: 50 bpm – 240 bpm
    • Resolution: 1 bpm
    • Accuracy: ±2 bpm
  • DECG:
    • Technique: Peak-to-peak detection
    • FHR Range: 50 bpm – 210 bpm (International Standard)
    • Resolution: 1 bpm
    • Accuracy: ±1 bpm
    • Input Impedance: > 10 M (Differential)
    • Input Impedance: > 20 M (Common Mode)
  • TOCO:
    • TOCO Range: 0 – 100 Relative (%)
    • Resolution: 1%
    • Zero Mode: Auto/Manual
    • Non-Linear Error: ±1 mmHg
  • IUP:
    • Pressure Range: 0 – 100 mmHg
    • Resolution: 1%
    • Zero Mode: Auto/Manual
    • Non-Linear Error: ±mmHg
  • Auto Fetal Movement (AFM):
    • Technique: Pulsed Doppler Ultrasound
    • Range: 0 – 100 (%)
    • Resolution: 1%
  • Marking: Manual fetal movement mark
  • SpO2:
    • Measuring Range: 0 – 100 Relative (%)
    • Resolution: 1%
    • Measuring Accuracy: ±2 bpm
  • NIBP:
    • Systolic Pressure: 50 mmHg – 240 mmHg
    • Mean Pressure: 25 mmHg – 200 mmHg
    • Diastolic Pressure: 15 mmHg – 180 mmHg
    • Resolution: 1 mmHg
  • Temperature:
    • Measuring Range: 28°C – 44°C
    • Accuracy: ±0.2°C